HDV-CDTP

Conductivity/Temperature Sensor
 
HDV-CDTP
Conductivity 0...30 mS/cm
Temperature 0...100 °C
6-pole electrode in carbon plastic technology
Different modes for temperature compensation
Galvanically isolated RS485 Interface
Complete ready to measure
OEM-Versions on request
Evaluation kit available



Ready to measure


The sensors has an galvanically isolated RS485-Interface and comes calibrated for immediate use. The sensor delivers measuring values for conductivity and temperature. The conductivity is temperature compensated.

Conductivity Measurement

The sensor has a state of the art 6-pole carbon plastic technology for conductivity measurement. A leakage current protection combined with grounding electrodes at both ends of the fluid path makes the sensor extremely insensitive to disturbances introduced in the fluid stream. This technology gives no polarization effects and results in an linear and stable function over years of use.

Conductivity Temperature Compensation

Conductivity temperature coefficient may be adjusted, ensuring a high accuracy conductivity measurement. The sensor offers four modes for compensation.

Temperature Compensation Modes
No compensation
  Linear with one adjustable value
 
Dynamic with two adjustable
values and temperatures
 
Non Linear Function for tap, purified
and waste water
in accordance with
ISO 7888 DIN EN 27888 Standards.

Evaluation Kit

We offer an evaluation kit which includes senor with an additional connector for USB-Connection, USB-Cable and the Biotech-Lab software. This software allows to display, record, save and print measurements.
 

 


 
 

    
   
 

Technical
Data
 
Conductivity   6 pole carbon plastic electrode.
4 measuring ranges.
Temperature compensation
coefficient may be adjusted.
Range: 0…30.00 mS/cm
Accuracy: 0…199 uS/cm 0,3% ± 1,5 uS/cm
  200…1999 uS/cm ± 6 uS/cm
  2...11.99 mS/cm ± 0.06 mS/cm
  12...16.00 mS/cm ± 0.03 mS/cm
  16...30 mS/cm ± 0.06 mS/cm
Temperature   Sensor integrated into the conductivity electrode
Range: 0...100 °C
Resolution: 0.01 °C
Accuracy: 25...40°C ± 0.1°C
otherwise ±  0.15°C 
Calibration   PTB/NIST traceable
Communication rate   All measurements five times per second
Material   In contact with measuring media:
PEI, Stainless Steel 1.4305, O-Ring VMQ
Flow path   5 mm inner diameter
Tube connection   Hose nozzle, outer diameter 7 mm or
Thread G1/8"
Interface   RS485, galvanically isolated
Power supply   3,3 to 5 V, max. 60 mA via RS485-Interface
Environmental
conditions
  Operating Temperature Range: 0 to 50°C
Storage Temperature Range: -40 to 80°C
Operating and Storage Humidity: 85% max
relative humidity non-condensing
from 0 to 50°C
Dimensions   133 x 26 x 44 mm
 
   
 
 
Home Produkte Unternehmen Support eMail Adresse
    HDC Biotech-Lab   Über uns   Verkauf & Service   Schreiben Sie uns   IBP Medical GmbH
Ikarusallee 15
30179 Hannover
Germany
+49 511 651647
From US 1-866-243-0187
 
    Dialyse Referenz Messgeräte   Qualitätsmanagement   FAQ   Ihre Stimme zählt  
    Druck & Durchfluss-Messgeräte   Neuigkeiten   Allgm. Geschäftsbedingungen   Newsletter  
    HDU-Sensoren              
    HDV-Sensoren       Garantie      
    Kalibrierservice              
    Kalibriersysteme              
    Hämodialysator Analyser              
    Kundenspezifische Entwicklung              

© Copyright 2015 IBP Medical GmbH   Zurück nach oben