HDU-PRH & HDU-PRS

Family of pressure sensors

HDU-PRH
High Accuracy |
 |
HDU-PRS |
 |
Pressure/Vacuum Sensor |
 |
 |
Pressure/Vacuum Sensor |
 |
Readings
in mmHg, kPa,
Bar or PSI |
 |
Readings
in mmHg, kPa,
Bar or PSI |
 |
Pressure Drop
Measurement |
 |
Pressure Drop
Measurement |
 |
Measures gasses |
 |
Measures gasses and fluids |
For
every application the right senor

Whatever your
requirements are, the HDU-Pressure sensors
fulfil the
need for high
accuracy and reliability. An intelligent product using
innovative and intelligent
technology, delivering great value.
For the following
applications we recommend.
 |
HDU-PRS30 for Hemodialysis
- resistance against dialysate |
 |
HDU-PRH15 for Blood
Pressure Calibration |
Technical Data
|
|
|
|
All Versions |
Reading: |
Readings as mmHg, kPa, Bar
or PSI |
Measuring: |
Temperature compensated |
Pressure Connection: |
IBP-Sensor_Port and G1/8" Female
internal thread |
Membrane |
HDU-PRS versions have a
Stainless steel membrane which is fluid resistant |
|
|
|
|
 |
|
Pressure
|
|
HDU-PRH5 |
Range: |
-5 to 5 psi
-0.35 to 0,35 bar
-259 to +259 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.05% full scale |
Media Compatibility: |
Gasses |
Over
pressure |
2 x
full scale |

HDU-PRH15 |
Range: |
-13 to 15 psi
-0.90 to 1 bar
-672 to +776 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.05% full scale |
Media Compatibility: |
Gasses |
Over
pressure |
2 x
full scale |

HDU-PRH30 |
Range: |
-13 to 30 psi
-0.90 to 2.07 bar
-672 to +1551 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.05% full scale |
Media Compatibility: |
Gasses |
Over
pressure |
1.5 x
full scale |

HDU-PRH50 |
Range: |
-13 to 50 psi
-0.90 to 3,45 bar
-672 to 2586 mmHg |
Resolution: |
0.1 mmHg |
Accuracy: |
0.05% full scale |
Media Compatibility: |
Gasses |
Over
pressure |
2 x
full scale |

HDU-PRH100 |
Range: |
-13 to 100 psi
-0.90 to 6,90 bar
-672 to 5172 mmHg |
Resolution: |
0.1 mmHg |
Accuracy: |
0.05% full scale |
Media Compatibility: |
Gasses |
Over
pressure |
2 x
full scale |
|
 |
HDU-PRS5 |
Range: |
-5 to 5 psi
-0.35 to 0,35 bar
-259 to +259 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.1% full scale |
Media Compatibility: |
Gasses and Fluids |
Over
pressure |
2 x
full scale |

HDU-PRS15 |
Range: |
-13 to 15 psi
-0.90 to 1 bar
-672 to +776 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.1% full scale |
Media Compatibility: |
Gasses and Fluids |
Over
pressure |
2 x
full scale |

HDU-PRS30 |
Range: |
-13 to 30 psi
-0.90 to 2.07 bar
-672 to +1551 mmHg |
Resolution: |
0.01 mmHg |
Accuracy: |
0.1% full scale
|
Media Compatibility: |
Gasses and Fluids |
Over
pressure |
1.5 x
full scale |

HDU-PRS100 |
Range: |
-13 to 100 psi
-0.90 to 6,90 bar
-672 to 5172 mmHg |
Resolution: |
0.1 mmHg |
Accuracy: |
0.1% full scale |
Media Compatibility: |
Gasses and Fluids |
Over
pressure |
2 x
full scale |
|
 |
 |
 |
|
|
Calibration |
|
PTB/NIST traceable |
|
|
 |
 |
 |
|
|
Communication
Interface |
|
RS485-Bus Fault Tolerant
115.200 kbps Transmission
Communication rate: Five
times per second |
|
|
 |
 |
 |
|
|
Degree of
protection |
|
IP67 |
|
|
 |
 |
 |
|
|
Power supply |
|
3,3 to 5 V, max. 60 mA
via RS485- or USB-Interface |
|
|
 |
 |
 |
|
|
Environmental
conditions |
|
Operating Temperature Range: 0 to 50°C
Storage Temperature Range: -40 to 80°C
Operating and Storage Humidity: 85% max
relative humidity non-condensing
from 0 to 50°C |
|
|
 |
 |
 |
|
|
Dimensions |
|
80 x 34 mm |
|
|
|
|