What is the calibration period of IBP sensors and meters?

The calibration period is one year. We recommend to double check the conductivity once per month with a
conductivity reference solution. If biological material grows inside the cell, it could interfere the accuracy
of the conductivity measurement.

We know that competing companies do not recommend a monthly test of the conductivity measurement,
some even offer this as a "feature" of their meters. From our more than 25 years experience in developing
dialysis meters we strongly recommend monthly testing as a safeguard precaution.
 


     
  If I send my sensor or meter in for calibration, what must I included in my shipment?

As the conductivity electrode is part of the calibration process, it must be included with the meter.
If your meter has a flow probe, it must also be included with your shipment. There is no need to
ship the battery charger or other accessories.
 


     
  Why do IBP Conductivity Sensors allow you to select the Temperature Coefficient of the Dialysis Machine to be calibrated?

This is a frequently misunderstood topic that is often neglected in hemodialysis.
Click here to read the background.
PDF (23 kByte)
 


 
   
 
               
  Home Products Company Support eMail us Address  
  SmartHDM About us Sales & Service Mail us IBP Medical GmbH
Ikarusallee 15
30179 Hannover
Germany
     
+49 511 957 336 0
 
From US 1-866-243-0187
 
 
    Dialysis Meters Privacy Statement FAQ We want your feedback
    Pressure & Flow Meters QM-Certificate Meter Pick Up Service Sign our Newsletter
    HDU-Sensors   Terms & Conditions  
    Calibration Service   Warranty  
    Calibration Systems   Worldwide Distributors  
        Online Invoice Payment  
 
 
 
  Copyright 2019 IBP Medical GmbH       back to top