IBP Medical
If quality matters - IBP Medical
 
 
 
Dialysis Reference Meters          
 
 
 
 
HDM97BQ 
HDM97BP
HDM97BO
HDM97BN

Reference Meters for the Service and Calibration of Hemodialysis Machines

 

HDM97BQ HDM97BP HDM97BO HDM97BN
Conductivity Conductivity Conductivity Conductivity
Temperature Temperature Temperature Temperature
Pressure Pressure Pressure    
Flow pH        

 

 

Innovative Details


Multifunctional

In using for hemodialysis, the meters can be used with any brand of dialysis machine.
Conductivity Measurement Years of experience with modern design and testing have proven that our 4-pole
electrode in carbon plastic technology provides superior performance.
This provides reliable, stable and linear results for conductivity measurement.

IBP is the only manufacturer of dialysis reference meters using 4-pole electrode
construction for conductivity measurement.

Also features facility for conductivity measurement of RO water.
The electrode is intrinsically tied to the meter and therefore cannot be misplaced.

Conductivity Temperature
Compensation

Conductivity temperature coefficients are selectable by brand name,
ensuring high accuracy conductivity measurement.
Pressure Measurement
 
Provides readings in mmHg, kPa Bar or PSI.
A special mode allows measuring of the pressure drop.
   
The HDM97BQ, HDM97BP and HDM97BO versions can be delivered with
a higher pressure measurement accuracy via an option, which allows
the calibration of blood pressure measurement devices in accordance with
DIN EN 1060 standard.
   
For the calibration of blood pressure measurement devices we offer a
calibration set according to EN1060.

For more details, please refer to
"How to calibrate blood pressure measuring devices"

Flow Measurement  
We provide an option for the
HDM97BQ that provides readings
in ml/min as well as showing the accumulated value.

More details on flow sensors.

pH Measurement Automatic temperature compensation by measuring or fixed temperature.
Stop Watch  
The stop watch function allows accurate time measurement
to 100 m/sec. resolution.
Field Calibration HDM97Pocket meters can quickly and easily be adjusted and calibrated in the field,
thus minimising loss of time.
USB-Interface USB 2.0 - electrically insulated in accordance with ISO 60601-1.
Data exchange with other software is possible via the USB-Interface.
See also Data Acquisition

Registration

Manufactured under certified quality management system in accordance to
ISO13485 standard.

Registered as a medical device in USA and Canada.
Sensor connectors  
HDM97BQ   HDM97BQ with Option C  (Connector for Conductivity probe)
 
HDM97BP HDM97BO
 

HDM97BN
   
   
     
Technical Data
 
   
Conductivity 4 measuring ranges - Quadropole electrode
Temperature compensation is user-selectable depending on the brand of dialysis machine
Range: 0 30,00 mS/cm
Accuracy: 0 200 uS/cm 0.6 uS/cm
201 2000 uS/cm 6 uS/cm
12 ... 16 mS/cm 0.03 mS/cm
otherwise 0.06 mS/cm

With option C, the conductivity/temperature probe may be disconnected.
See picture above.
This option is available for all meters.
 

Temperature Sensor integrated in the conductivity electrode
Range: 0...100 C
Resolution: 0.01 C
Accuracy: 25 ... 40C 0.05 C, otherwise   0.1 C 
Pressure  Reading selectable as mmHg, kPa, Bar or PSI

Version HDM97BQ, BP and BO

Range:   -700 to + 1900 mmHg, 2,5 Bar
Resolution:   0.1 mmHg
Accuracy:   0 ... 300 mmHg 1 mmHg
otherwise 2 mmHg
    in temperature range 20 to 23 C
Ceramic sensor fluid resistant

Option H with higher accuracy
Range:   -700 to + 1900 mmHg, 2,5 Bar
Resolution:   0.01 mmHg
Accuracy:   0 ... 300 mmHg 0.5 mmHg,
301 ... 1000 mmHg 1 mmHg
otherwise 2 mmHg
    in temperature range 20 to 23 C
Sensor not fluid resistent
Flow    Only Version BQ
Version 100.2000  
Measuring principal: Turbine
Range:   100 to 2000 ml/min
H2O at  22C / 71.6F
Resolution:  1 ml/min
Accuracy: 2.0 % of measured value
Repeatability better 0,5%
Fluid temperature 0C to +80C
Pressure 0 to 5 bar at 22C
Version 5.1750
High Accuracy
 
Measuring principal: Gear
Range: 5 to 1750 ml/min
H2O at 22C / 71.6F
Resolution:  0,01 ml/min
Accuracy: 1.5 % of measured value
Repeatability better 0,5%
Fluid temperature 0C to +80C
Pressure -0.1 to 5 bar at 22C
pH       Only Version BP Temperature via temperature probe in conductivity cell or default value
Range: 0...14 pH
Resolution: 0.01 pH
Accuracy: 0.02 pH
Power Supply Li-Ion-Polymer battery, rechargeable with 5 V DC, 500 mA
External charger:
Input Voltage 100 245 V / 50/60 Hz
Output Voltage 5 V DC, 500 mA
Or via USB-Interface from a computer
Operates up to 200 hours with fully charged battery
Dimensions and Weight 165 x 82 x 33 mm - 6.5" x 3.2" x 1.3
0.4 Kg - 0.88 lbs including probe
 
   
 
               
  Home Products Company Support eMail us Address  
  Biotech-Lab About us Sales & Service Mail us IBP Medical GmbH
Ikarusallee 15
30179 Hannover
Germany
     
+49 511 651647
 
From US 1-866-243-0187
 
 
    Dialysis Meters Privacy Statement FAQ We want your feedback
    Pressure & Flow Meters QM-Certificate Meter Pick Up Service Sign our Newsletter
    Simulator and Tester News Terms & Conditions  
    HDU-Sensors   Warranty  
    HDV-Sensors   Worldwide Distributors  
    Calibration Service   Online Invoice Payment  
    Calibration Systems      
    Dialyser Analyser      
    Customer-Specific Development      
 
 
 
  Copyright 2018 IBP Medical GmbH       back to top