IBP Medical
If quality matters - IBP Medical
 
 
 
Hemodialyser Analyser          
 
 
 
 
Most pictures
can
be zoomed.


 

 

Technical Data
Water preparation    
Delivery   600 ml/min
Degassing   Combination of heat and
negative pressure
Air content   less than 9,8 % O2
Dialyser    
Measuring fluid
Blood and
dialysate side
  Range 100 500 ml/min
Resolution 1 ml/min
Accuracy +/- 1,5 %

Temperature

  Range 35 40C
Accuracy +/- 0.3C
Pressure at dialyser Blood and
dialysate side
each in and out
  Range -700 3420 mmHg
Accuracy 0 600 mmHg
              +/- 1 mmHg
Otherwise +/- 2 mmHg
UFR-Measurement   Range 5 500 ml/min
Resolution 0.5 ml/min
Accuracy +/- 1,0 %
of measured value
Interfaces   USB 2.0
LAN
Power Supply   230 V 50/60 Hz   3500 W
other upon request
Dimensions    145 x 80 x 66 cm H x W x D
Weight   92 Kg
     

 
   
 
               
  Home Products Company Support eMail us Address  
  Biotech-Lab About us Sales & Service Mail us IBP Medical GmbH
Ikarusallee 15
30179 Hannover
Germany
     
+49 511 651647
 
From US 1-866-243-0187
 
 
    Dialysis Meters Privacy Statement FAQ We want your feedback
    Pressure & Flow Meters QM-Certificate Meter Pick Up Service Sign our Newsletter
    HDU-Sensors News Terms & Conditions  
    Calibration Service   Warranty  
    Calibration Systems   Worldwide Distributors  
        Online Invoice Payment  
 
 
 
  Copyright 2018 IBP Medical GmbH       back to top